Four core parameters. One matched set, confirmed in writing before it ships.
Tell us the parameter, the measuring point and the controller you already use. We configure the sensor, the controller and the required accessories as one matched set — and confirm signal output, wiring and mounting in writing before the order is placed. Every order is calibrated and documented before shipment, with calibration and pre-shipment test records supplied.
| Parameter | Sensor Type | Output Signal | Matched Controller | Mounting | What We Verify Before Dispatch |
|---|---|---|---|---|---|
| pH / ORP | Analog glass pH or ORP electrode (mV); digital RS485 pH/ORP sensor with built-in PT1000 temperature compensation | mV (analog) · 4–20 mA via transmitter · RS485 Modbus RTU | Single-parameter pH/ORP controller, or multi-parameter controller with 4–20 mA or Modbus input, alarm relays and optional retransmission | Immersion · inline flow cell · insertion | Analog mV sensors need a nearby transmitter for long cable runs; digital sensors require a matching Modbus register map. Temperature compensation and cable length are checked against the controller input. |
| Residual Chlorine | Amperometric membrane-covered three-electrode residual chlorine sensor; reagent-based DPD colorimetric analyzer where compliance monitoring is required | 4–20 mA · RS485 Modbus RTU | Residual chlorine controller or analyzer with pH compensation input, flow detection and dosing relays | Constant-flow flow cell · panel-mounted analyzer | The HOCl/OCl⁻ reading is pH-dependent, so pH compensation and stable flow are mandatory. Reagent consumables, membrane and electrolyte replacement intervals are stated up front. |
| Turbidity | 90° scattered-light nephelometric sensor to ISO 7027, inline or immersion, with optional automatic wiper | 4–20 mA · RS485 Modbus RTU | Turbidity controller with measuring-range selection, wiper control and alarm relays | Inline pipe · immersion · flow-through cell | Measuring range in NTU, wiper requirement for fouling water, and formazin calibration standard are confirmed. Stray-light and bubble effects are checked against the installation. |
| Dissolved Oxygen | Optical luminescent DO sensor, or polarographic Clark-type sensor | RS485 Modbus RTU · 4–20 mA with built-in temperature compensation | DO controller with temperature, salinity and barometric compensation, plus aeration control relays | Immersion · insertion · flow cell | Optical versus polarographic input compatibility, compensation parameters, cable length and calibration method (air-saturated water / zero-oxygen) are confirmed in writing. |
For full product listings, specifications and model codes, see the Products section. Other parameters are supported on a project basis.
For each measuring point we specify the sensor, the controller or transmitter, and the accessories — flow cells, mounting brackets, cables and calibration kits — as one compatible set. Calibration and pre-shipment test records and batch traceability ship with every order, with one point of contact from quotation to after-sales.
These four cards jump to the matching row in the matrix above. They are not product categories.
Request a Written Compatibility Confirmation Contact Us